Absolute point dose measurement of stereotactic radiotherapy plan for ultra-small tumors
RADIOBIOLOGY AND RADIOMEDICINE|更新时间:2025-02-28
|
Absolute point dose measurement of stereotactic radiotherapy plan for ultra-small tumors
“In the field of stereotactic radiotherapy for ultra small volume tumors, experts have proposed a real-time cross calibration method for ionization chambers, which achieves accurate measurement of absolute point dose and provides a solution for improving treatment accuracy.”
Journal of Radiation Research and Radiation ProcessingVol. 39, Issue 6, Pages: 69-76(2021)
作者机构:
浙江大学医学院附属第一医院放射治疗科 杭州 310003
作者简介:
CHEN Yuanhua (male) was born in April 1987, and obtained his master's degree from Nanjing University of Aeronautics and Astronautics in 2015. Now he is a medical physicist at the Department of Radiotherapy of the First Affiliated Hospital, Zhejiang University School of Medicine
LU Zhongjie, chief technician, E-mail: luzhongjie@zju.edu.cn
CHEN Yuanhua,LU Zhongjie.Absolute point dose measurement of stereotactic radiotherapy plan for ultra-small tumors[J].Journal of Radiation Research and Radiation Processing,2021,39(06):69-76.
CHEN Yuanhua,LU Zhongjie.Absolute point dose measurement of stereotactic radiotherapy plan for ultra-small tumors[J].Journal of Radiation Research and Radiation Processing,2021,39(06):69-76. DOI: 10.11889/j.1000-3436.2021.rrj.39.060302.
Absolute point dose measurement of stereotactic radiotherapy plan for ultra-small tumors