1.上海市计量测试技术研究院 上海 201203
[ "陆逊, 男, 1977年11月出生, 2014年于上海理工大学毕业, 医疗器械工程专业, 1999年就职于上海市计量测试技术研究院, 计量检测岗位, 工程师, 从事电离辐射与医学计量, E-mail:lux@simt.com.cn", "LU Xun (male) was born in November 1977, and graduated from University of Shanghai for Science and Technology in 2014, majoring in medical equipment engineering. Now he engages in ionizing radiation and medical metrology in Shanghai Institute of Measurement and Testing Technology from 1999. E-mail:lux@simt.com.cn" ]
眭国平, 高级工程师, E-mail:suigp@simt.com.cnSUI Guoping, senior engineer, E-mail:suigp@simt.com.cn
扫 描 看 全 文
陆逊, 眭国平, 杭雪莲, 等. 医用电子加速器高能光子水吸收剂量溯源[J]. 辐射研究与辐射工艺学报, 2017,35(6):060701-4.
Xun LU, Guoping SUI, Xuelian HANG, et al. Traceability of absorbed dose to water in high-energy photon beams in medical electron accelerators[J]. Journal of Radiation Research and Radiation Processing, 2017,35(6):060701-4.
陆逊, 眭国平, 杭雪莲, 等. 医用电子加速器高能光子水吸收剂量溯源[J]. 辐射研究与辐射工艺学报, 2017,35(6):060701-4. DOI: 10.11889/j.1000-3436.2017.rrj.35.060701.
Xun LU, Guoping SUI, Xuelian HANG, et al. Traceability of absorbed dose to water in high-energy photon beams in medical electron accelerators[J]. Journal of Radiation Research and Radiation Processing, 2017,35(6):060701-4. DOI: 10.11889/j.1000-3436.2017.rrj.35.060701.
采用直接法测量医用电子加速器高能光子水吸收剂量,与我国现行的从空气比释动能转化为水吸收剂量间接法测量的结果进行比较。结果表明,两种方法测量结果相差1%以内,同时用传递比较法进行了验证。证明我国基于加速器光子水吸收剂量基准而建设的水吸收剂量溯源体系是可行的,为我国水吸收剂量直接测量提供依据。
Results obtained by direct measurement of absorbed dose to water for medical electron accelerators were compared with that by indirect measurements (through converting the measured air kerma into absorbed dose to water). Experiments results demonstrate that the difference between two methods is within 1%, it is validated by the transfer comparison method meanwhile. It is demonstrated that the absorbed dose to water traceability system based on the photon absorbed dose to water for the accelerator is feasible. This provides the basis for direct measurement of absorbed dose to water for accelerators.
高能光子水吸收剂量辐射质修正因子量值传递量值溯源
High energy photonAbsorbed dose to waterCorrection factor for beam qualityDissemination of the value of quantityTraceability of value of quantity
全国电离辐射计量技术委员会. JJF1035-2006电离辐射计量术语及定义[S]. 北京: 中国计量出版社, 2007.
National Ionizing Radiation Metrology Technical Committee. JJF1035-2006 metrological terms and their definitions for ionizing radiation[S]. Beijing:China Metrology Publishing House, 2007.
全国电离辐射计量技术委员会. JJG589-2008医用电子加速器辐射源检定规程[S]. 北京: 中国计量出版社, 2009.
National Ionizing Radiation Metrology Technical Committee. JJG589-2008 verification regulation of medical electron accelerator radiation source[S]. Beijing:China Metrology Publishing House, 2009.
International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy:an international code of practice for dosimetry based on standards of absorbed dose to water, technical report series No.398[S]. Viena, IAEA, 2000.
张 圆月 , 杨 勇 , 黄 扬 , 等 . 医用高能光子水吸收剂量测量规范TRS-277与TRS-398比较 . 中国测试技术 , 2009 . 35 (2 ):14 -17 . http://kns.cnki.net/KCMS/detail/detail.aspx?filename=sycs200902006&dbname=CJFD&dbcode=CJFQhttp://kns.cnki.net/KCMS/detail/detail.aspx?filename=sycs200902006&dbname=CJFD&dbcode=CJFQ, .
Yuanyue ZHANG , yong YANG , Yang HUANG , 等 . Comparison between TRS-277 and TRS-398:absorbed dose to water for clinical high-energy photon beams . China Measurement & Testing Technology , 2009 . 35 (2 ):14 -17 . http://kns.cnki.net/KCMS/detail/detail.aspx?filename=sycs200902006&dbname=CJFD&dbcode=CJFQhttp://kns.cnki.net/KCMS/detail/detail.aspx?filename=sycs200902006&dbname=CJFD&dbcode=CJFQ, .
全国法制计量管理技术委员会. JJF1033-2016计量标准考核规范[S]. 北京: 中国质检出版社, 2016.
National Legal Metrology Management Technical Committee. JJF1033-2016 rule for the examination of measurement standards[S]. Beijing:China Quality Inspection Publishing House, 2016.