顾思毅, 王孝娃, 袁宝峰, 等. 低剂量条件下调制传递函数法与线对法的相关性分析[J]. 辐射研究与辐射工艺学报, 2015,33(3):30201-030201.
GU Siyi, WANG Xiaowa, YUAN Baofeng, et al. Correlation analysis for modulation transfer function method and line pair method under loe dose conditions[J]. Journal of Radiation Research and Radiation Processing, 2015,33(3):30201-030201.
顾思毅, 王孝娃, 袁宝峰, 等. 低剂量条件下调制传递函数法与线对法的相关性分析[J]. 辐射研究与辐射工艺学报, 2015,33(3):30201-030201. DOI: 10.11889/j.1000-3436.2015.rrj.33.030201.
GU Siyi, WANG Xiaowa, YUAN Baofeng, et al. Correlation analysis for modulation transfer function method and line pair method under loe dose conditions[J]. Journal of Radiation Research and Radiation Processing, 2015,33(3):30201-030201. DOI: 10.11889/j.1000-3436.2015.rrj.33.030201.
通过在低剂量条件下对两种CT图像空间分辨率获得方法(调制传递函数法与线对法)的相关性分析,提供CT质量控制过程中空间分辨率评价的客观依据。研究使用GE Light Speed 16 CT对位于扫描正中线的Catphan 500体模中的CPT528及CPT486模块进行扫描。扫描条件:标准胸部扫描条件(120kV、200mA)以及低剂量扫描条件(电压:80kV、100kV、120kV、和140kV 4档;电流:10-60mA,10mA递增)。各扫描条件均使用CPT486模块测量噪声。调节窗宽窗位,观察不同扫描条件下可分辨的线对(为说明主观性对线对法的影响,随机选取部分CPT528扫描图像,由5名受试者进行读数,进行统计分析)。测量不同扫描图像中的点扩散函数(Point spread function,PSF),再对PSF进行二维傅里叶变换得到调制传递函数(Modulation transfer function,MTF),最后使用SAS8.2软件对10%MTF值与线对法目测值进行统计分析,寻找两种空间分辨率获得方法(MTF法与线对法)之间的相关性。结果表明,当MTF为10%时对应的空间分辨率与线对法目测值一致(t’=0.18、p=0.8556)。低剂量条件下部分MTF曲线受噪声影响导致MTF值增大。
The aim of this study is to analyze the correlation of two spatial resolution of CT image acquisition method(modulation transfer function method and line pair method) under low dose conditions,which to provide an objective basis for spatial resolution evaluation of CT quality control.This research scanned located mid-line CPT 528 and 486 module of Catphan 500 phantom by using GE Light Speed 16 CT under standard chest scanning conditions(120kV,200mA) and low dose conditions(80kV,100kV,120kV,and 140kV four gears; 10-60mA,10mA increased) separately,with CPT486 module measuring noise.Adjust the window wide and window level,observe distinguishable lines on the card under different scanning conditions(to illustrate the impact of subjectivity on the line pairs method,randomly selected parts of CPT528 scanned images were read by five testers and the results were statistically analyzed).The point spread function(PSF) on the different locations were measured via the program,and then on two-dimensional Fourier transformation of PSF to get modulation transfer function(MTF).Finally,we used SAS8.2 software statistical to analyze 10% MTF values and visual values in order to search the correlation between MTF method and line pairs method.The results show that when the MTF is 10% the corresponding spatial resolution and visual values obtained by lines pair method are consistent(t’=0.18,p=0.8556).Under low-dose conditions,part of MTF curves are influenced by noise,leading to increase of MTF value.
低剂量胸部CT调制传递函数(MTF)空间分辨率噪声
Low-dose chest CTModulation transfer function(MTF)Spatial resolutionNoise