THE MODEL BPS TE SPHERICAL CAVITY IONIZATION CHAMBER
|更新时间:2023-04-23
|
THE MODEL BPS TE SPHERICAL CAVITY IONIZATION CHAMBER
Journal of Radiation Research and Radiation ProcessingVol. 2, Issue 2, Pages: 52-58(1984)
作者机构:
1.中国科学院生物物理研究所
作者简介:
基金信息:
DOI:
CLC:
扫 描 看 全 文
Zhang Zhonglun, Li Fengzhang. THE MODEL BPS TE SPHERICAL CAVITY IONIZATION CHAMBER. [J]. Journal of Radiation Research and Radiation Processing 2(2):52-58(1984)
DOI:
Zhang Zhonglun, Li Fengzhang. THE MODEL BPS TE SPHERICAL CAVITY IONIZATION CHAMBER. [J]. Journal of Radiation Research and Radiation Processing 2(2):52-58(1984)DOI:
THE MODEL BPS TE SPHERICAL CAVITY IONIZATION CHAMBER
The model BPS TE spherical chambers have been constructed. The basic performances of the chambers have been determined. The chambers are three terminal devices with gas inlet and outlet During operating the chamber at 250 V and 500 V, the saturation correction factors are 1.003 and 1.0007, respectively. The relative change of the response for the chamber is less than 0.5 per cent when the angles of the incident radiation are changed in the range of 0°~120° for the ,60,Co gamma rays. The stem effect and the energy response of the chamber have been studied. The results of measurement show that the variation of the composition of TE-gas will cause noticeable error