STUDY FOR POTENTIOMETRIC ESTIMATION OF ROUTINE HIGH DOSE WITH THE CERIC-CEROUS DOSIMETER
|更新时间:2023-04-23
|
STUDY FOR POTENTIOMETRIC ESTIMATION OF ROUTINE HIGH DOSE WITH THE CERIC-CEROUS DOSIMETER
Journal of Radiation Research and Radiation ProcessingVol. 14, Issue 1, Pages: 24-28(1996)
作者机构:
1.江苏省农科院原子能利用研究所,南京 210014
作者简介:
基金信息:
DOI:
CLC:
扫 描 看 全 文
Li Naining, Zhou Yunsheng, Fang Bingbing, et al. STUDY FOR POTENTIOMETRIC ESTIMATION OF ROUTINE HIGH DOSE WITH THE CERIC-CEROUS DOSIMETER. [J]. Journal of Radiation Research and Radiation Processing 14(1):24-28(1996)
DOI:
Li Naining, Zhou Yunsheng, Fang Bingbing, et al. STUDY FOR POTENTIOMETRIC ESTIMATION OF ROUTINE HIGH DOSE WITH THE CERIC-CEROUS DOSIMETER. [J]. Journal of Radiation Research and Radiation Processing 14(1):24-28(1996)DOI:
STUDY FOR POTENTIOMETRIC ESTIMATION OF ROUTINE HIGH DOSE WITH THE CERIC-CEROUS DOSIMETER
The method for potentiometric estimation of megarad dose with the ceric-cerous dosimetry system has been discussed in this paper. The difference in electrochemical potential between irradiated and unirradiated solution was used for the measurement of absorbed dose instead of the complex spectrophotometric procedure. Results for the doses ranging from 10 to 20 kGy are presented that the data obtain from the potentiometry with single distilled water and container without special cleaning are agreement within 4.5% with that from the Fricke system. Since electrodes and a pH meter used in the method are very popular in a common chemical laboratory and dilution is not required on the ceric-cerous dosimeter solution, the potentiometry of the ceric-cerous system is very suitable for routine dose measurement on radiation processing.