Analysis on dosimetric verification of flattening filter free conventional segmentation partial arc RapidArc plans
RADIOBIOLOGY AND RADIOMEDICINE|更新时间:2025-02-28
|
Analysis on dosimetric verification of flattening filter free conventional segmentation partial arc RapidArc plans
“In the field of radiation therapy, researchers compared the application of two detectors in the validation of the partial arc RapidArc program and found that the aSi-1200 EPID has a higher gamma pass rate, but both detectors cannot reliably identify MLC beam model errors, providing a reference for explaining the high failure rate of IROC-H phantom experiments.”
Journal of Radiation Research and Radiation ProcessingVol. 39, Issue 6, Pages: 77-86(2021)