郎淑玉, 张仲纶. 适用于辐射加工的谷氨酰胺晶溶发光剂量计及其BPLL型测量装置[J]. 辐射研究与辐射工艺学报, 1986,4(1):35-41.
Lang Shuyu, Zhang Zhonglun. THE GIUTAMINE LYOLUMINESCENCE DOSIMETER APPLICABLE TO RADIATION PROCESSING AND MODEL BPLL READOUT EQUIPMENT[J]. Journal of Radiation Research and Radiation Processing, 1986,4(1):35-41.
本文介绍了我们实验室研制的BPLL型晶溶发光剂量计的测量装置。本装置采用直流电流积分法。选用GDB-52型光电倍增管做光探测器,使用NWF-02 A型微电流放大器和电压—频率转换器。装置的灵敏度可进行调节,对于剂量测量,八小时的稳定性是令人满意的。48个测量数据的平均值的标准偏差为±0.4%。在此装置上对适用于辐射加工的谷氨酰胺晶溶发光剂量计的剂量学特性进行了研究。谷氨酰胺的剂量响应线性范围为10,2,~10,4,Gy。论述了溶剂的温度、样品的质量对光产额的影响以及射线照射后的存贮效应。
The model BPLL readout equipment of lyoluminescence dosimeter has been constructed in our lab. The D. C. current integration method has been employed in the equipment. The model GDB-52 photomultiplier has been selected as a light detector, the model NWF-02 A weak current amplifier and a voltage-frequency converter have been used. The sensitivity of the equipment can be regulated and its stability in 8 hours is satisfactory for dose measurement, the standard deviation of the mean of 48 measured values is ±0.4%. The glutamine lyoluminescence dosimetric characteristics for radiation processing has been studied on the model BPLL readout equipment. The range of linear dose response is in the 10,2,~10,4, Gy. The dependence of light yield on the temperature of solvent and the mass of samples and the storage effect of samples after irradiation are disscused.
晶溶发光、电流积分法谷氨酰胺测量装置剂量学特性
lyoluminescencecurrent integration methodglutaminereadout equipmentdosimetric characteristics